A comparison of headnet electrode arrays for electrical impedance tomography of the human head
Article
Tidswell, A., Bagshaw, A., Holder, D., Yerworth, R., Eadie, L., Murray, S., Morgan, L. and Bayford, R. 2003. A comparison of headnet electrode arrays for electrical impedance tomography of the human head. Physiological Measurement. 24 (2), pp. 527-544. https://doi.org/10.1088/0967-3334/24/2/363
Type | Article |
---|---|
Title | A comparison of headnet electrode arrays for electrical impedance tomography of the human head |
Authors | Tidswell, A., Bagshaw, A., Holder, D., Yerworth, R., Eadie, L., Murray, S., Morgan, L. and Bayford, R. |
Abstract | Three types of commercially available headnet electrode arrays, designed for use in EEG, and conventional EEG Ag/AgCl cup electrodes were tested on human subjects, and a realistic, saline-filled head-shaped tank was prepared with vegetable skin to simulate human skin in order to determine the optimum electrode system for electrical impedance tomography (EIT) of the human head. Impedance changes during EIT acquisition were produced in healthy volunteers during a finger–thumb apposition task and in tanks by the insertion of a Perspex rod. Signal-to-baseline noise, measured from raw EIT data, was 2.3 ± 0.3 and 2.3 ± 0.2 for the human and tank data, respectively. In both the human and tank experiments, a commercial hydrogel elasticated electrode headnet produced the least amount of baseline noise, and was the only headnet in the human data with noise levels acceptable for EIT imaging. Image quality measured in the tank was similar for most of the headnets tested, except that the EEG electrodes produced a higher positional error and electrodes in a geodesic elasticated net produced images with worse subjective image quality. Overall, the hydrogel elasticated headnet was judged to be the most suitable for human neuroimaging with EIT. |
Research Group | Biophysics and Bioengineering group |
Publisher | Institute of Physics. |
Journal | Physiological Measurement |
ISSN | 0967-3334 |
Publication dates | |
30 Apr 2003 | |
Publication process dates | |
Deposited | 26 May 2009 |
Output status | Published |
Digital Object Identifier (DOI) | https://doi.org/10.1088/0967-3334/24/2/363 |
Language | English |
https://repository.mdx.ac.uk/item/81q6w
51
total views0
total downloads3
views this month0
downloads this month
Export as
Related outputs
Clinical utility of ultrasound imaging for measuring anterior thigh thickness after anterior cruciate ligament injury in an individual patient to assess postsurgery outcome
Mechelli, F., Bayford, R., Garelick, H., Stokes, M. and Agyapong-Badu, S. 2023. Clinical utility of ultrasound imaging for measuring anterior thigh thickness after anterior cruciate ligament injury in an individual patient to assess postsurgery outcome. Case Reports in Orthopedics. 2023. https://doi.org/10.1155/2023/6672951Thermodynamics of mechanopeptide sidechains
Haque, M., Kadir, M. and Bayford, R. 2023. Thermodynamics of mechanopeptide sidechains. AIP Advances. 13 (8). https://doi.org/10.1063/5.0154129Effects of patient recumbency position on neonatal chest EIT
Seifnaraghi, N., De Gelidi, S., Frerichs, I., Kallio, M., Sorantin, M., Demosthenous, A. and Bayford, R. 2023. Effects of patient recumbency position on neonatal chest EIT. IEEE Access. 11, pp. 68257 - 68268. https://doi.org/10.1109/ACCESS.2023.3290904Advances in electrical impedance tomography and bioimpedance including applications in COVID-19 diagnosis and treatment
Bayford, R., Rosalind, S. and Frerichs, I. 2022. Advances in electrical impedance tomography and bioimpedance including applications in COVID-19 diagnosis and treatment. Physiological Measurement. 43 (2). https://doi.org/10.1088/1361-6579/ac4e6cEffect of routine suction on lung aeration in critically ill neonates and young infants measured with electrical impedance tomography
Händel, C., Becher, T., Miedema, M., Kallio, M., Papadouri, T., Waldmann, A.D., Sophocleous, L., Yerworth, R., Bayford, R., Rimensberger, P.C., van Kaam, A.H. and Frerichs, I. 2023. Effect of routine suction on lung aeration in critically ill neonates and young infants measured with electrical impedance tomography. Scientific Reports. 13 (1). https://doi.org/10.1038/s41598-023-42965-7An 89.3% current efficiency, sub 0.1% THD current driver for electrical impedance tomography
Li, J., Jiang, D., Wu, Y, Neshatvar, N., Bayford, R. and Demosthenous, A. 2023. An 89.3% current efficiency, sub 0.1% THD current driver for electrical impedance tomography. IEEE Transactions on Circuits and Systems II: Express Briefs. 70 (10), pp. 3742-3746. https://doi.org/10.1109/TCSII.2023.3294753Bronchodilator effect on regional lung function in pediatric viral lower respiratory tract infections
Strodthoff, C., Kähkönen, T., Bayford, R., Becher, T., Frerichs, I. and Kallio, M. 2022. Bronchodilator effect on regional lung function in pediatric viral lower respiratory tract infections. Physiological Measurement. 43 (10). https://doi.org/10.1088/1361-6579/ac9450High frame rate electrical impedance tomography system for monitoring of regional lung ventilation
Rahal, M., Dai, J., Wu, Y., Bardill, A., Bayford, R. and Demosthenous, A. 2022. High frame rate electrical impedance tomography system for monitoring of regional lung ventilation. 44th Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC). Glasgow, Scotland, United Kingdom 11 - 15 Jul 2022 IEEE. pp. 2487-2490 https://doi.org/10.1109/EMBC48229.2022.9871479Development of a biosensor for fast point-of-care blood analysis of Troponin
Bayford, R., Damaso, R., Jiang, D., Rahal, M. and Demosthenous, A. 2022. Development of a biosensor for fast point-of-care blood analysis of Troponin. 44th Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC). Glasgow, Scotland, United Kingdom 11 - 15 Jul 2022 IEEE. pp. 910-913 https://doi.org/10.1109/EMBC48229.2022.9871851A low-power recursive I/Q signal generator and current driver for bioimpedance applications
Hanzaee, F., Neshatvar, N., Rahal, M., Jiang, D., Bayford, R. and Demosthenous, A. 2022. A low-power recursive I/Q signal generator and current driver for bioimpedance applications. IEEE Transactions on Circuits and Systems II: Express Briefs. 69 (10), pp. 4108-4112. https://doi.org/10.1109/TCSII.2022.3187076![](/~17/aqs/default-thumbnail.png)