The effect of serial data collection on the accuracy of electrical impedance tomography images
Article
Yerworth, R. and Bayford, R. 2013. The effect of serial data collection on the accuracy of electrical impedance tomography images. Physiological Measurement. 34 (6), pp. 659-669. https://doi.org/10.1088/0967-3334/34/6/659
Type | Article |
---|---|
Title | The effect of serial data collection on the accuracy of electrical impedance tomography images |
Authors | Yerworth, R. and Bayford, R. |
Abstract | There has been a surge of interest in using electrical impedance tomography (EIT) for monitoring regional lung ventilation, however, EIT is an ill-conditioned problem, and errors/noise in the boundary voltages can have an undesirable effect on the quality of the final image. Most EIT systems in clinical usage use serial data collection hence data used to create a single image will have been collected at different times. This paper presents a study of the resulting image distortion, and proposes a method for correcting this lag in situations where the frame rate is insufficient to prevent significant image degradation. Significant correlation between the standard deviation of the time dependent reciprocity error and time delay dLe between the reciprocal electrode combinations was found for both adult and neonate data. This was reduced when the data was corrected for dLe. Original and corrected data was reconstructed with the GREIT algorithm and visible differences were found for the neonate data. Ideally EIT systems should be run at a frame rate of at least 50 times the frequency of the dominant and interesting physiological signals. Where this is not practical, the intra-frame system timings should be determined and lag corrected for. |
Research Group | Biophysics and Bioengineering group |
Publisher | IOP Publishing |
Journal | Physiological Measurement |
ISSN | 0967-3334 |
Electronic | 1361-6579 |
Publication dates | |
29 May 2013 | |
Publication process dates | |
Deposited | 28 Apr 2015 |
Accepted | 07 May 2013 |
Output status | Published |
Digital Object Identifier (DOI) | https://doi.org/10.1088/0967-3334/34/6/659 |
Language | English |
https://repository.mdx.ac.uk/item/852x4
48
total views0
total downloads6
views this month0
downloads this month
Export as
Related outputs
Clinical utility of ultrasound imaging for measuring anterior thigh thickness after anterior cruciate ligament injury in an individual patient to assess postsurgery outcome
Mechelli, F., Bayford, R., Garelick, H., Stokes, M. and Agyapong-Badu, S. 2023. Clinical utility of ultrasound imaging for measuring anterior thigh thickness after anterior cruciate ligament injury in an individual patient to assess postsurgery outcome. Case Reports in Orthopedics. 2023. https://doi.org/10.1155/2023/6672951Thermodynamics of mechanopeptide sidechains
Haque, M., Kadir, M. and Bayford, R. 2023. Thermodynamics of mechanopeptide sidechains. AIP Advances. 13 (8). https://doi.org/10.1063/5.0154129Effects of patient recumbency position on neonatal chest EIT
Seifnaraghi, N., De Gelidi, S., Frerichs, I., Kallio, M., Sorantin, M., Demosthenous, A. and Bayford, R. 2023. Effects of patient recumbency position on neonatal chest EIT. IEEE Access. 11, pp. 68257 - 68268. https://doi.org/10.1109/ACCESS.2023.3290904Advances in electrical impedance tomography and bioimpedance including applications in COVID-19 diagnosis and treatment
Bayford, R., Rosalind, S. and Frerichs, I. 2022. Advances in electrical impedance tomography and bioimpedance including applications in COVID-19 diagnosis and treatment. Physiological Measurement. 43 (2). https://doi.org/10.1088/1361-6579/ac4e6cEffect of routine suction on lung aeration in critically ill neonates and young infants measured with electrical impedance tomography
Händel, C., Becher, T., Miedema, M., Kallio, M., Papadouri, T., Waldmann, A.D., Sophocleous, L., Yerworth, R., Bayford, R., Rimensberger, P.C., van Kaam, A.H. and Frerichs, I. 2023. Effect of routine suction on lung aeration in critically ill neonates and young infants measured with electrical impedance tomography. Scientific Reports. 13 (1). https://doi.org/10.1038/s41598-023-42965-7An 89.3% current efficiency, sub 0.1% THD current driver for electrical impedance tomography
Li, J., Jiang, D., Wu, Y, Neshatvar, N., Bayford, R. and Demosthenous, A. 2023. An 89.3% current efficiency, sub 0.1% THD current driver for electrical impedance tomography. IEEE Transactions on Circuits and Systems II: Express Briefs. 70 (10), pp. 3742-3746. https://doi.org/10.1109/TCSII.2023.3294753Bronchodilator effect on regional lung function in pediatric viral lower respiratory tract infections
Strodthoff, C., Kähkönen, T., Bayford, R., Becher, T., Frerichs, I. and Kallio, M. 2022. Bronchodilator effect on regional lung function in pediatric viral lower respiratory tract infections. Physiological Measurement. 43 (10). https://doi.org/10.1088/1361-6579/ac9450High frame rate electrical impedance tomography system for monitoring of regional lung ventilation
Rahal, M., Dai, J., Wu, Y., Bardill, A., Bayford, R. and Demosthenous, A. 2022. High frame rate electrical impedance tomography system for monitoring of regional lung ventilation. 44th Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC). Glasgow, Scotland, United Kingdom 11 - 15 Jul 2022 IEEE. pp. 2487-2490 https://doi.org/10.1109/EMBC48229.2022.9871479Development of a biosensor for fast point-of-care blood analysis of Troponin
Bayford, R., Damaso, R., Jiang, D., Rahal, M. and Demosthenous, A. 2022. Development of a biosensor for fast point-of-care blood analysis of Troponin. 44th Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC). Glasgow, Scotland, United Kingdom 11 - 15 Jul 2022 IEEE. pp. 910-913 https://doi.org/10.1109/EMBC48229.2022.9871851A low-power recursive I/Q signal generator and current driver for bioimpedance applications
Hanzaee, F., Neshatvar, N., Rahal, M., Jiang, D., Bayford, R. and Demosthenous, A. 2022. A low-power recursive I/Q signal generator and current driver for bioimpedance applications. IEEE Transactions on Circuits and Systems II: Express Briefs. 69 (10), pp. 4108-4112. https://doi.org/10.1109/TCSII.2022.3187076![](/~17/aqs/default-thumbnail.png)