Effect of sternal electrode gap and belt rotation on the robustness of pulmonary electrical impedance tomography parameters
Article
Sophocleous, L., Waldmann, A. D., Becher, T., Kallio, M., Rahtu, M., Miedema, M., Papadouri, T., Karaoli, C., Tingay, D. G., Van Kaam, A. H., Yerworth, R., Bayford, R. and Frerichs, I 2020. Effect of sternal electrode gap and belt rotation on the robustness of pulmonary electrical impedance tomography parameters. Physiological Measurement. 41 (3). https://doi.org/10.1088/1361-6579/ab7b42
Type | Article |
---|---|
Title | Effect of sternal electrode gap and belt rotation on the robustness of pulmonary electrical impedance tomography parameters |
Authors | Sophocleous, L., Waldmann, A. D., Becher, T., Kallio, M., Rahtu, M., Miedema, M., Papadouri, T., Karaoli, C., Tingay, D. G., Van Kaam, A. H., Yerworth, R., Bayford, R. and Frerichs, I |
Abstract | Objective: Non-adhesive textile electrode belts offer several advantages over adhesive electrodes and are increasingly used in neonatal patients during continuous electrical impedance tomography (EIT) lung monitoring. However, non-adhesive belts may rotate in unsedated patients and discrepancies between chest circumference and belt sizes may result in a gap between electrodes near the sternum. This project aimed to determine the effects of belt rotation and sternal electrode gap on commonly used lung EIT parameters. Approach: We developed a simulation framework based on a 3D finite-element model and introduced lung regions with little or no ventilation that could be changed according to a decremental positive end-expiratory pressure (PEEP) trial. Four degrees of sternal gap and belt rotation were simulated and their effect on the EIT parameters silent spaces, centre of ventilation, global inhomogeneity index and overdistension/collapsed lung (OD/CL) analysed. Additionally, seven premature infants were examined to assess the influence of leftward and rightward belt rotations in a clinical setting. Main results: Small violations of the electrode equidistance criterion and rotations of the belts less than one electrode space exert only minor effects on the EIT parameters and do not impede the interpretation. Rotations of two and three electrode spaces induce non-negligible effects that might lead to flawed interpretations. The 'best PEEP' determined with the OD/CL approach was robust and identifiable with all studied sternal gaps and belt rotations. Significance: We revealed an important challenge for neonatal EIT applications related to a wide electrode gap at the sternum and belt rotation, which should be avoided in clinical application. |
Keywords | EIT; electrical impedance; respiratory system; electrode rotation; electrode spacing; neonatal lung imaging; EIT belt; electrodes; positive end-expiratory pressure |
Publisher | IOP Publishing |
Journal | Physiological Measurement |
ISSN | 0967-3334 |
Electronic | 1361-6579 |
Publication dates | |
Mar 2020 | |
Online | 17 Apr 2020 |
Publication process dates | |
Deposited | 24 Aug 2023 |
Accepted | 28 Feb 2020 |
Submitted | 22 Oct 2019 |
Output status | Published |
Digital Object Identifier (DOI) | https://doi.org/10.1088/1361-6579/ab7b42 |
Web of Science identifier | WOS:000528530300001 |
Language | English |
https://repository.mdx.ac.uk/item/8zv19
69
total views0
total downloads1
views this month0
downloads this month
Export as
Related outputs
Chest EIT based on Lagrange Multipliers reconstruction
Seifnaraghi, N. and Bayford, R. 2024. Chest EIT based on Lagrange Multipliers reconstruction. Xie, L., Fu, F., Long, Y., Ge, H., Pan, Q. and Zhao, Z. (ed.) 24th International Conference on Biomedical Applications of Electrical Impedance Tomography. Hangzhou, China 27 - 30 Jun 2024A 1.76 mW, 355-fps, electrical impedance tomography system with a simple time-to-digital impedance readout for fast neonatal lung imaging
Li, J., Jiang, D., Wu, Y., Zhang, J., Seifnaraghi, N., Bayford, R. and Demosthenous, A. 2024. A 1.76 mW, 355-fps, electrical impedance tomography system with a simple time-to-digital impedance readout for fast neonatal lung imaging. IEEE Journal of Solid-State Circuits. https://doi.org/10.1109/JSSC.2024.3434638Progress in electrical impedance tomography and bioimpedance
Bayford, R., Sadleir, R., Frerichs, I., Oh, T. and Leonhardt, S. 2024. Progress in electrical impedance tomography and bioimpedance. Physiological Measurement. 45 (8). https://doi.org/10.1088/1361-6579/ad68c1A current DAC based current generator with fourth-order current-mode filter for electrical impedance tomography
Li, J., Wu, Y., Jiang, D., Bayford, R. and Demosthenous, A. 2024. A current DAC based current generator with fourth-order current-mode filter for electrical impedance tomography. 2024 IEEE International Symposium on Circuits and Systems. Singapore, Singapore 19 - 22 May 2024 IEEE. pp. 1-4 https://doi.org/10.1109/iscas58744.2024.10558679Clinical utility of ultrasound imaging for measuring anterior thigh thickness after anterior cruciate ligament injury in an individual patient to assess postsurgery outcome
Mechelli, F., Bayford, R., Garelick, H., Stokes, M. and Agyapong-Badu, S. 2023. Clinical utility of ultrasound imaging for measuring anterior thigh thickness after anterior cruciate ligament injury in an individual patient to assess postsurgery outcome. Case Reports in Orthopedics. 2023. https://doi.org/10.1155/2023/6672951Live demonstration: real time imaging with electrical impedance tomography
Zohoori, S., Rahal, M., Habibollahi, M., Jiang, D., Wu, Y., Bardill, A., Seifnaraghi, N., Yerworth, R., Bayford, R. and Demosthenous, A. 2023. Live demonstration: real time imaging with electrical impedance tomography. 2023 IEEE Biomedical Circuits and Systems Conference (BioCAS). Toronto, Canada 19 - 21 Oct 2023 IEEE. https://doi.org/10.1109/BioCAS58349.2023.10388836Live demonstration: a bioimpedance-based robotic hand control platform using a customised neural network
Yao, T., Almarri, N., Wu, Y., Jiang, D., Bayford, R. and Demosthenous, A. 2023. Live demonstration: a bioimpedance-based robotic hand control platform using a customised neural network. 2023 IEEE Biomedical Circuits and Systems Conference (BioCAS). Toronto, Canada 19 - 21 Oct 2023 IEEE. https://doi.org/10.1109/biocas58349.2023.10389001A compact neural network for high accuracy bioimpedance-based hand gesture recognition
Yao, T., Wu, Y., Jiang, D., Bayford, R. and Demosthenous, A. 2023. A compact neural network for high accuracy bioimpedance-based hand gesture recognition. 2023 IEEE Biomedical Circuits and Systems Conference (BioCAS). Toronto, Canada 19 - 21 Oct 2023 IEEE. https://doi.org/10.1109/biocas58349.2023.10388679Effect of routine suction on lung aeration in critically ill neonates and young infants measured with electrical impedance tomography
Händel, C., Becher, T., Miedema, M., Kallio, M., Papadouri, T., Waldmann, A.D., Sophocleous, L., Yerworth, R., Bayford, R., Rimensberger, P.C., van Kaam, A.H. and Frerichs, I. 2023. Effect of routine suction on lung aeration in critically ill neonates and young infants measured with electrical impedance tomography. Scientific Reports. 13 (1). https://doi.org/10.1038/s41598-023-42965-7An 89.3% current efficiency, sub 0.1% THD current driver for electrical impedance tomography
Li, J., Jiang, D., Wu, Y, Neshatvar, N., Bayford, R. and Demosthenous, A. 2023. An 89.3% current efficiency, sub 0.1% THD current driver for electrical impedance tomography. IEEE Transactions on Circuits and Systems II: Express Briefs. 70 (10), pp. 3742-3746. https://doi.org/10.1109/TCSII.2023.3294753Thermodynamics of mechanopeptide sidechains
Haque, M., Kadir, M. and Bayford, R. 2023. Thermodynamics of mechanopeptide sidechains. AIP Advances. 13 (8). https://doi.org/10.1063/5.0154129Effects of patient recumbency position on neonatal chest EIT
Seifnaraghi, N., De Gelidi, S., Frerichs, I., Kallio, M., Sorantin, M., Demosthenous, A. and Bayford, R. 2023. Effects of patient recumbency position on neonatal chest EIT. IEEE Access. 11, pp. 68257 - 68268. https://doi.org/10.1109/ACCESS.2023.3290904Prolonged continuous monitoring of regional lung function in infants with respiratory failure
Becher, T.H., Miedema, M., Kallio, M., Papadouri, T., Karaoli, C., Sophocleous, L., Rahtu, M., Van Leuteren, R.W., Waldmann, A.D., Strodthoff, C., Yerworth, R., Dupré, A., Benissa, M.-R., Nordebo, S., Khodadad, D., Bayford, R., Vliegenthart, R., Rimensberger, P.C., Van Kaam, A.H. and Frerichs, I. 2022. Prolonged continuous monitoring of regional lung function in infants with respiratory failure . Annals of the American Thoracic Society. 19 (6), pp. 873-1080. https://doi.org/10.1513/AnnalsATS.202005-562OCA low power, low THD current driver with discrete common-mode feedback for EIT applications
Li, J., Wu, Y., Bayford, R., Jiang, D. and Demosthenous, A. 2022. A low power, low THD current driver with discrete common-mode feedback for EIT applications . 29th IEEE International Conference on Electronics, Circuits and Systems. Glasgow, UK 24 - 26 Oct 2022 IEEE. https://doi.org/10.1109/ICECS202256217.2022.9971123Generation of anatomically inspired human airway tree using electrical impedance tomography: A method to estimate regional lung filling characteristics
Zamani, M., Kallio, M., Bayford, R. and Demosthenous, A. 2022. Generation of anatomically inspired human airway tree using electrical impedance tomography: A method to estimate regional lung filling characteristics. IEEE Transactions on Medical Imaging. 41 (5), pp. 1125-1137. https://doi.org/10.1109/TMI.2021.3136434Advances in electrical impedance tomography and bioimpedance including applications in COVID-19 diagnosis and treatment
Bayford, R., Rosalind, S. and Frerichs, I. 2022. Advances in electrical impedance tomography and bioimpedance including applications in COVID-19 diagnosis and treatment. Physiological Measurement. 43 (2). https://doi.org/10.1088/1361-6579/ac4e6cA low-power recursive I/Q signal generator and current driver for bioimpedance applications
Hanzaee, F., Neshatvar, N., Rahal, M., Jiang, D., Bayford, R. and Demosthenous, A. 2022. A low-power recursive I/Q signal generator and current driver for bioimpedance applications. IEEE Transactions on Circuits and Systems II: Express Briefs. 69 (10), pp. 4108-4112. https://doi.org/10.1109/TCSII.2022.3187076High frame rate electrical impedance tomography system for monitoring of regional lung ventilation
Rahal, M., Dai, J., Wu, Y., Bardill, A., Bayford, R. and Demosthenous, A. 2022. High frame rate electrical impedance tomography system for monitoring of regional lung ventilation. 44th Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC). Glasgow, Scotland, United Kingdom 11 - 15 Jul 2022 IEEE. pp. 2487-2490 https://doi.org/10.1109/EMBC48229.2022.9871479Development of a biosensor for fast point-of-care blood analysis of Troponin
Bayford, R., Damaso, R., Jiang, D., Rahal, M. and Demosthenous, A. 2022. Development of a biosensor for fast point-of-care blood analysis of Troponin. 44th Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC). Glasgow, Scotland, United Kingdom 11 - 15 Jul 2022 IEEE. pp. 910-913 https://doi.org/10.1109/EMBC48229.2022.9871851Bronchodilator effect on regional lung function in pediatric viral lower respiratory tract infections
Strodthoff, C., Kähkönen, T., Bayford, R., Becher, T., Frerichs, I. and Kallio, M. 2022. Bronchodilator effect on regional lung function in pediatric viral lower respiratory tract infections. Physiological Measurement. 43 (10). https://doi.org/10.1088/1361-6579/ac9450![](/~24/bgc/default-thumbnail.png)